色综合久久中文字幕无码_日韩欧美在线国产精品_成年男人看片免费视频播放人_国语免费一级毛片私人_羞羞AV手机在线观看_heyzo无码综合国产精品_紧缚波多野结衣_穆斯林妇女大BBw_99re热这里只有精品视频中文字幕_欧美天堂视频

7*24小時(shí)工作時(shí)間隨時(shí)全心服務(wù)

0531-85919987

產(chǎn)品導(dǎo)航

Technology

您的當(dāng)前位置:首頁(yè) > Technology
Marine heavy oil freezing point automatic determination solution
2020-12-10 09:14:31 admin 441
Marine heavy oil refers to the low fat fuel oil generally refers to the low quality, difficult to use, low price Marine fuel (commonly known as heavy oil), the heavy oil used by Marine diesel engines is mostly "intermediate fuel oil" or "residual oil", its viscosity, poor can be called heavy fuel oil.
The maximum temperature when the fuel is cooled to the liquid level in the test condition is called the freezing point. When the fuel temperature is lower than the freezing point, it cannot be pumped.
Determination of Marine heavy oil condensate is carried out in accordance with the Petroleum Products Condensate Method GB/T 510-1983 (1991).
Suitable for determination of freezing point of dark oil products and lubricating oil.
And SH113E1 automatic condensation point detector based on national standard national standard GB/T510 based on petroleum products solidifying point measurement, adopt modern advanced technology, mechanical, optical, electronic and computer technology into an organic whole, can automatically complete the condensation point of petroleum products testing, the instrument adopts photoelectric detection technology, automatic inclined glass tube method, imported compressor refrigeration system to ensure that meet the requirements of refrigeration and depth.
Apparatus needed for experiment
1.SH113E1 automatic freezing point analyzer, developed and produced by Shandong Shengtai Instrument Co., LTD.
2. Coagulation point test tube in accordance with GB/T510.
 
Experimental steps:
Preparation work
1. After unpacking the instrument, please count the accessories according to the packing list. If there is any damage, please contact our company.
2. The instrument should be handled vertically. The tilt Angle should not exceed 30 degrees along the height of the instrument.
3. Connect the power supply of the instrument to the socket with a capacity of 220V/10A or above, and pay attention to having a good grounding wire.
Ii. Experiment process
1. Freezing point determination: put the rubber plug cover with sensor and detector on the test tube, and put the freezing point test tube into the freezing point cold filter point bath.
2. Click the freezing point measurement to enter the freezing point measurement interface.

3. The main interface of freezing point determination is shown in the figure above.
The above four buttons are respectively the main interface, parameter setting, system setting and historical data.
The main display area is the trend chart of the temperature change of the sample over time, which can show the temperature value of the sample and the metal bath. The display area on the right will show the specific value, and below is the expected freezing point and test tube test results.
In the lower right corner are the start button, tilt button and close button.
4. After all the preparation work is finished, click the start button to carry out the test. During the test, the system will automatically detect the freezing point; after the test, the buzzer will prompt and print the freezing point result.